BINDER MKT240-480V-C Series MKT - Dynamic Climate Chambers, For Rapid Temperature Changes With Extended Low Temperature Range, 9020-0364
Average cooling down time according to IEC 60068-3-5 [K/min]: 4.2
Average heating-up rate according to IEC 60068-3-5 [K/min]: 5
Cooling down time from 180 °C to -70 °C [min]: 95
Max. heat compensation at 25 °C [W]: 3200
Temperature fluctuation depending on setpoint [± K]: 0.1…0.4
Temperature range: -70…180
Temperature variation depending on setpoint [± K]: 0.1…1
Benefits
Homogeneous climate conditions thanks to APT.line™ technology
Comprehensive programming and data acquisition
Large heated viewing window
Important characteristics
Temperature range: -70 °C to +180 °C
4 zero-voltage relay contacts
APT.line™ preheating chamber technology
Programmable condensation protection for test material
Heated viewing window with LED interior lighting
BINDER Multi Management software APT-COM™ Basic Edition
Troubleshooting system with visual and audible alarms
Intuitive touchscreen controller with time-segment and real-time programming
Internal data logger, measured values can be read out in open format via USB
Access port with silicone plug: 50 mm, left
Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
4 castors, two with brakes
Computer interface: Ethernet
Adjustable ramp function
Integrated chart recorder
Real-time clock
Door heating
1 stainless steel rack
Inner chamber made of stainless steel
CFC-free refrigerants R-452A and R-23
Cooling with cascade compressor cooling unit
Technical specifications
Performance Data Temperature:
Average cooling down time according to IEC 60068-3-5 [K/min]: 4.2
Average heating-up rate according to IEC 60068-3-5 [K/min]: 5
Cooling down time from 180 °C to -70 °C [min]: 95
Max. heat compensation at 25 °C [W]: 3200
Temperature fluctuation depending on setpoint [± K]: 0.1…0.4
Temperature range: -70…180
Temperature variation depending on setpoint [± K]: 0.1…1
Electrical data
Nominal power [kW]: 6.5
Phase (Nominal voltage): 3~
Power frequency [Hz]: 60
Rated Voltage [V]: 480
Unit fuse [A]: 16
Doors
Unit doors: 1
Housing dimensions not incl. fittings and connections
Depth net [in]: 36.45
Height net [in]: 76.2
Width net [in]: 43.9
Internal Dimensions
Depth [in]: 17.45
Height [in]: 27.58
Width [in]: 28.96
Measures
Interior volume [cu.ft.]: 8
Load per rack [lbs]: 66
Net weight of the unit (empty) [lbs]: 648
Permitted load [lbs]: 154
Viewing window height [in]: 11.82
Viewing window width [in]: 20.02
Wall clearance back [in]: 11.82
Wall clearance sidewise [in]: 7.88
Environment-specific data
Sound-pressure level [dB(A)]: 67
Fixtures
Number of shelves (std./max.): 1/6
| Brand | BINDER |
|---|---|
| UNSPSC | 41103410 |