BINDER CB056UL-120V-O Series CB - CO2 Incubators, With Hot Air Sterilization And Heat Sterilizable Co2 Sensor, 9640-0008
Quick Overview
Recovery time after 30 seconds door open at 37 °C [min]: 5 Temperature fluctuation at 37 °C [± K]: 0.1 Temperature range +6 °C above ambient temperature to: 60 °C Temperature variation at 37 °C [± K]: 0.3
Details
Benefits
Safe – cell cultures receive maximum protection against contamination Reliable – constant well-being conditions for the cells Smart – simple routine cleaning and convenient operation Economical – efficient operation without consumables
Important characteristics
Optimized humidification system with condensation protection Tried-and-tested anti-contamination concept with 180 °C hot air sterilization Hot-air sterilizable CO₂ sensor with single-beam infrared technology Seamless inner chamber made of stainless steel with flanges as shelf support system Fanless interior with Venturi CO₂ gas mixing nozzle Intuitive touchscreen controller Internal data recording, can be read out in open format via USB interface Fault diagnosis system with visual and acoustic alarm Interfaces: Ethernet, USB, zero-voltage alarm contact Fail-safe CO₂ system for protecting the pH of cell cultures
Optional equipment
O₂ control with O₂ control ranges: 0.2 to 20 vol.% O₂ or 10 to 95 vol.% O₂ Multi-sectional glass doors with special shelves for dividing up the space to accommodate different experiments Stacking adapter for secure, space-saving installation Gas tank changer for connecting two gas tanks to one or two incubators
Technical specifications
Performance Data Temperature
Recovery time after 30 seconds door open at 37 °C [min]: 5 Temperature fluctuation at 37 °C [± K]: 0.1 Temperature range +6 °C above ambient temperature to: 60 °C Temperature variation at 37 °C [± K]: 0.3
Performance Data Climate
Humidity range: 90…95
CO₂-Data
CO₂ measurement IR CO₂ range [Vol.-% CO₂] 0…20 CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]: 5
O₂-Data
O₂ control ranges with option: O₂ range [Vol.-% O₂]: 10…95 O₂ Recovery time after 30 seconds door open at 10 Vol.-% O₂ [min]: 5 O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]: 5 Standard O₂ control ranges: O₂ range [Vol.-% O₂]: 0.2…20
Electrical data
Nominal power [kW]: 0.8 Phase (Nominal voltage): 1~ Power frequency [Hz]: 50/60 Rated Voltage [V]: 100…120 Unit fuse [A]: 16
Doors
Inner doors: 1 Unit doors: 1
Housing dimensions not incl. fittings and connections
Depth net [in]: 21.47 Height net [in]: 26.56 Width net [in]: 20.8