BINDER MKFT115-400V Series MKFT - Dynamic Climate Chambers, For Rapid Temperature Changes With Humidity Control And Extended Low Temperature Range, 9020-0382
Quick Overview
Average cooling down time according to IEC 60068-3-5 [K/min]: 4.2 Average heating-up rate according to IEC 60068-3-5 [K/min]: 5.5 Max. heat compensation at 25 °C [W]: 1500 Temperature fluctuation depending on setpoint [± K]: 0.1…0.1 Temperature range: -70…180 Temperature variation depending on setpoint [± K]: 0.1…1.3
Details
Benefits
Homogeneous climate conditions thanks to APT.line™ technology Automatic water and wastewater management Pressure humidification with fast response times Comprehensive programming and data acquisition Large heated viewing window
Important characteristics
Temperature range: -70 °C to +180 °C Humidity range: 10 % to 98 % RH Integrated water-storage tank, 20 liters 4 zero-voltage relay contacts APT.line™ preheating chamber technology Programmable condensation protection for test material Intuitive touchscreen controller with time-segment and real-time programming Internal data logger, measured values can be read out in open format via USB Heated viewing window with LED interior lighting Humidity regulation with capacitative humidity sensor and vapor humidification BINDER Multi Management Software APT-COM™ Basic Edition Troubleshooting system with visual and audible alarms Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm Access port with silicone plug: 50 mm, left 4 castors, two with brakes Computer interface: Ethernet Adjustable ramp function Integrated chart recorder Real-time clock Door heating Alarm notification in the event of insufficient water in fresh water tank 1 stainless steel rack Complete safety connection kit for water supply and drainage, up to 1 m in height Inner chamber made of stainless steel CFC-free refrigerants R-452A and R-23 Cooling with cascade compressor cooling unit
Technical specifications Performance Data Temperature:
Average cooling down time according to IEC 60068-3-5 [K/min]: 4.2 Average heating-up rate according to IEC 60068-3-5 [K/min]: 5.5 Max. heat compensation at 25 °C [W]: 1500 Temperature fluctuation depending on setpoint [± K]: 0.1…0.1 Temperature range: -70…180 Temperature variation depending on setpoint [± K]: 0.1…1.3
Performance Data Climate
Dew point temperature range [°C]: 5…94 Humidity fluctuation depending on setpoint: ≤2,5 ± % RH Humidity range: 10…98 Temperature fluctuation depending on setpoint [± K]: 0.1…1 Temperature range [°C]: 10…95
Electrical data
Nominal power [kW]: 5 Phase (Nominal voltage): 3~ Power frequency [Hz]: 50 Rated Voltage [V]: 400 Unit fuse [A]: 16
Doors
Unit doors: 1
Housing dimensions not incl. fittings and connections
Depth net [in]: 34.08 Height net [in]: 68 Width net [in]: 38.61