BINDER BD400UL-120V Series BD Classic.Line - Standard-Incubators, With Gravity Convection, 9010-0176
Heating-up time to 37 °C [min]: 120
Recovery time after 30 seconds door open at 37 °C [min]: 10
Temperature fluctuation at 37 °C [± K]: 0.1
Temperature range +5 °C above ambient temperature to [°C]: 100
Temperature variation at 37 °C [± K]: 0.5
Benefits
Safe and reproducible incubation even under high batch throughputs in long-term operation
Disinfection routine at 100 °C
Stable atmosphere by inner glass door
Important characteristics
Temperature range: room temperature plus 5 °C to 100 °C
APT.line™ preheating chamber technology
gravity convection
Adjustable exhaust air flap
Controller with timer functions
Inner door made of tempered safety glass
2 chrome-plated racks
Class 3.1 independent temperature safety device (DIN 12880) with visual temperature alarm
Computer interface: RS 422
Technical specifications
Performance Data Temperature:
Heating-up time to 37 °C [min]: 120
Recovery time after 30 seconds door open at 37 °C [min]: 10
Temperature fluctuation at 37 °C [± K]: 0.1
Temperature range +5 °C above ambient temperature to [°C]: 100
Temperature variation at 37 °C [± K]: 0.5
Electrical data
Nominal power [kW]: 0.85
Phase (Nominal voltage): 1~
Power frequency [Hz]: 60
Rated Voltage [V]: 120
Unit fuse [A]: 12.5
Doors
Unit doors: 2
Housing dimensions not incl. fittings and connections
Depth net [in]: 30.14
Height net [in]: 40.4
Width net [in]: 48.7
Internal Dimensions
Depth [in]: 19.7
Height [in]: 31.52
Width [in]: 39.4
Measures
Interior volume [cu.ft.]: 14.1
Load per rack [lbs]: 77
Net weight of the unit (empty) [lbs]: 298
Permitted load [lbs]: 198
Wall clearance back [in]: 6.3
Wall clearance sidewise [in]: 3.94
Environment-specific data
Energy consumption at 37 °C [Wh/h]: 56
Fixtures
Number of shelves (std./max.): 2/9
| Brand | BINDER |
|---|---|
| UNSPSC | 41104400 |