Mfg Part Number
9020-0358

BINDER MKF240-480V-C Series MKF - Dynamic Climate Chambers, For Rapid Temperature Changes With Humidity Control, 9020-0358

Quick Overview

Average cooling down time according to IEC 60068-3-5 [K/min]: 5
Average heating-up rate according to IEC 60068-3-5 [K/min]: 5
Cooling down time from 180 °C to -40 °C [min]: 120
Max. heat compensation at 25 °C [W]: 2800
Temperature fluctuation depending on setpoint [± K]: 0.1…0.5
Temperature range: -40…180
Temperature variation depending on setpoint [± K]: 0.1…1.5

$46,970.98
Order will ship within 1-3 business days
Add to Wish List






Benefits

Homogeneous climate conditions thanks to APT.line™ technology
Automatic water and wastewater management
Pressure humidification with fast response times
Comprehensive programming and data acquisition
Large heated viewing window

Important characteristics

Temperature range: -40 °C to +180 °C
Humidity range: 10% to 98% RH
Humidity regulation with capacitive humidity sensor and steam humidification
APT.line™ preheating chamber technology
Intuitive touchscreen controller with time-segment and real-time programming
CFC-free refrigerant R-452A
Integrated water-storage tank, 20 L
1 stainless steel rack
Access port with silicone plug, 50 mm, left
4 zero-voltage relay contacts

Optional equipment

APT-COM Multi Management Software – for managing, recording, and documenting unit parameters
Racks – option of stainless steel racks, depending on test requirements
Systems for water treatment and compressed air supply
Services – wide range of services to ensure that the unit functions correctly

Technical specifications

Performance Data Temperature:

Average cooling down time according to IEC 60068-3-5 [K/min]: 5
Average heating-up rate according to IEC 60068-3-5 [K/min]: 5
Cooling down time from 180 °C to -40 °C [min]: 120
Max. heat compensation at 25 °C [W]: 2800
Temperature fluctuation depending on setpoint [± K]: 0.1…0.5
Temperature range: -40…180
Temperature variation depending on setpoint [± K]: 0.1…1.5

Performance Data Climate

Dew point temperature range [°C]: 5…94
Humidity fluctuation depending on setpoint: ≤2,5 ± % RH
Humidity range: 10…98
Temperature fluctuation depending on setpoint [± K]: 0.1…1.3
Temperature range [°C]: 10…95

Electrical data

Nominal power [kW]: 6.8
Phase (Nominal voltage): 3~
Power frequency [Hz]: 60
Rated Voltage [V]: 480
Unit fuse [A]: 16

Doors

Unit doors: 1

Housing dimensions not incl. fittings and connections

Depth net [in]: 36.45
Height net [in]: 67.6
Width net [in]: 43.9

Internal Dimensions

Depth [in]: 17.45
Height [in]: 27.58
Width [in]: 28.96

Measures

Interior volume [cu.ft.]: 8
Load per rack [lbs]: 66
Net weight of the unit (empty) [lbs]: 593
Permitted load [lbs]: 154
Viewing window height [in]: 11.82
Viewing window width [in]: 20.02
Wall clearance back [in]: 11.82
Wall clearance sidewise [in]: 7.88

Environment-specific data

Energy consumption at 25 °C and 60 % RH [Wh/h]: 1500
Sound-pressure level [dB(A)]: 67

Fixtures

Number of shelves (std./max.): 1/6

More Information
Brand BINDER
UNSPSC 41103418